CAUTION Risk of injury to the patient's eye!
• When switching from the peristaltic pump to the
Venturi or rotary pump during surgery without
changing settings, there is a Risk of injury to the
patient.
INFORMATION: the flow rate and the vacuum limit can be
set independently for the peristaltic pump, for Venturi
pump only vacuum can be set. When using the Venturi
pump, the current flow rate increases with vacuum
settings and aspiration cannula dimensions.
WARNING Risk of injury to the patient's eye!
By changing the height of the IV pole by forcing it
manually, it is possible to obtain an incorrect indication
of the height of the bottle and injure the patient.
• Do not force the IV pole to change its height.
CAUTION Risk of injury to the patient's eye!
It is possible to cause anterior chamber instability if the
height of the IV pole or the vacuum settings is not
correct!