R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
86
102
"Vacuum level error"
The vacuum exceeds the
preset value in Venturi
mode
Contact the technical
service of OPTIKON
2000 S.p.A.
103
"Invalid diathermy
power"
Fault in the diathermy
system (the output power
is much higher or lower
than the preset power)
Contact the technical
service of OPTIKON
2000 S.p.A.
104
"Ultrasound enabling
error"
Error in the ultrasonic
system (the output power
is not activated even if
the opposite is shown, due
to an activation port error
or a false contact of the
cables)
Contact the technical
service of OPTIKON
2000 S.p.A.
200 Error messages
Id
Message
Possible cause
Correction
measurement
200
"Air enabling error"
Error in the air infusion
system and controlled
irrigation (the secondary
transistor of the safety
switch is checked at
startup)
Contact the
technical service of
OPTIKON 2000 S.p.A.
201
"Watchdog error"
Error in the monitoring
circuit
(verified at startup)
Contact the
technical service of
OPTIKON 2000 S.p.A.
202
"Serial communication
error"
Communication error
between the graphical
user interface and the
control panel
Contact the
technical service of
OPTIKON 2000 S.p.A.
203
"Warning: active
diathermy!"
Fault in the diathermy
system (output power is
activated due to an
activation port error)
Contact the
technical service of
OPTIKON 2000 S.p.A.