R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
87
Id
Message
Possible cause
Correction
measurement
205
"Diathermy enabling
error"
Fault in the diathermy
system (the output power
is not activated even
though the opposite is
shown, due to an error in
the activation port or a
false contact of the
cables)
Contact the
technical service of
OPTIKON 2000 S.p.A.
206
"Reference voltage or
power supply error"
Incorrect reference
voltage in A/D transducer
or +5V power value
incorrect (verified at
startup)
Contact the
technical service of
OPTIKON 2000 S.p.A.
208
"Warning: active
ultrasound!"
Error in the ultrasound
system (output power is
activated due to an
activation port error)
Contact the
technical service of
OPTIKON 2000 S.p.A.
209
"Vacuum offset error"
Error during preparation:
excessive reading of the
sensor in the absence of a
vacuum
a) The preparation
procedure was
restarted after Err
210. Open the
infusion clamp on
the infusion unit,
briefly open the
continuous irrigation
and restart the
preparation
procedure.
b) The vacuum
sensor must be
calibrated. Contact
the OPTIKON 2000
S.p.A. technical
service to calibrate
the unit.