R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
88
Id
Message
Possible cause
Correction
measurement
210
"No irrigation"
Error during preparation:
The vacuum does not
decrease after opening
the venting valve.
Open the infusion
clamp on the
infusion unit, briefly
open the continuous
irrigation and restart
the preparation
procedure.
211
"Infusion pressure too
high"
The IV pole is located at
too high a level
Lower the IV pole as
far as possible
212
"IV pole error"
IV pole error
Contact the
technical service of
OPTIKON 2000 S.p.A.
213
216
"LED 1 over-heating"
"LED 2 over-heating"
The system for
illumination 1 or 2 has
reached an excessive
temperature.
Make sure the
ventilation grids on
the rear panel are
clear of obstructions.
Turn off the
superheated LED and
use one of the other
two light sources.
If the problem
persists, contact the
OPTIKON 2000 S.p.A.
technical service.