R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
89
300 Error messages
Id
Message
Possible cause
Correction
measurement
300
"Install the I/A
cassette"
The I/A cassette is not
connected
Connect the I/A
cassette
301
"I/A cassette full,
aspiration stopped"
The I/A cassette
collection tank is full
Replace the I/A
cassette
302
"Low input pressure"
The pressure of the
external air supply is too
low or equal to zero
a) The device is not
connected to the
compressed air supply
or the stopcock is
closed.
Connect the device of
the compressed air
source or open the
stopcock.
b) The compressed air
supply pressure of the
device is less than 450
kPa (65 psi).
Check the pressure in
the compressed air
circuit.
303
"Major leakage"
Error during preparation
(Priming): the peristaltic
pump can not establish a
vacuum of 100 mmHg
within 30 sec.
a) The test chamber is
not installed on the
sleeve. Ensure that
the test chamber is
properly installed on
the sleeve.
b) Irrigation/aspiration
tubes are not
connected to the
handpiece. Connect
the
irrigation/aspiration
tubings from the
cassette to the
handpiece.
c) The I/A cassette is
faulty.
Replace the I/A
cassette