R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
90
Id
Message
Possible cause
Correction
measurement
d) The pump system
must be recalibrated,
contact the OPTIKON
2000 S.p.A. technical
service.
304
"Aspiration line
occluded"
Error during preparation
(Priming): the vacuum
level during preparation
exceeds 300 mmHg
a) The handpiece and
tip have not been
properly cleaned
before sterilisation.
Replace the tip and/or
handpiece.
b) The I/A cassette is
faulty.
Replace the I/A
cassette
305
"Minor leakage"
Error during preparation
(Priming): the peristaltic
pump can not establish a
vacuum of 300mmHg
a) Irrigation/aspiration
connectors are not
fully inserted into the
handpiece connectors.
Connect the
connectors correctly.
b) The sleeve or test
chamber is not
mounted correctly.
Verify that the test
chamber is fully
engaged with the
sleeve and that the
sleeve is correctly
positioned on the
handpiece.
306
"Check the US tip"
Error during Priming: the
phaco handpiece can not
be tuned
a) The phaco tip is
loose. Tighten the
phaco tip in the
handpiece correctly
using a phaco tip
wrench.
b) The phaco tip is
damaged.
Inspect the phaco tip
and replace it as
necessary.