R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
91
Id
Message
Possible cause
Correction
measurement
c) the phaco
handpiece is damaged.
Replace the phaco
handpiece
307
"Weak phaco
handpiece"
The phaco handpiece can
not oscillate to more than
50 micrometres.
a) Check to see if the
phaco tip is loose, and
if so, tighten it
properly using a phaco
tip wrench.
b) It is normal for the
piezo-ceramic element
to deteriorate with
use and after a
number of sterilisation
cycles. Send the
handpiece to the
OPTIKON 2000
S.p.A.technical
service.
308
Install controlled
irrigation
The controlled irrigation
air line is not installed.
Install the controlled
infusion system before
continuing, or use the
gravity system.
310
"Low power US"
The ultrasound handpiece
can not oscillate with the
amplitude set
a) Check to see if the
phaco tip is loose, and
if so, tighten it
properly using a phaco
tip wrench.
b) It is normal for the
piezo-ceramic element
to deteriorate with
use and after a
number of sterilisation