R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
92
Id
Message
Possible cause
Correction
measurement
cycles. Return the
handpiece to the
OPTIKON 2000 S.p.A.
technical service.
311
Insert U/S handpiece
The U/S handpiece is not
connected to the unit
Connect the U/S
handpiece to the unit.
If the problem persists
even after, replace
the handpiece.
312
Please, prime
The handpiece is
connected but not
initialized
See "5.4.3 Installing
the phaco
accessories".
313
Please, prime (of the
vitrectomy cutter)
The vitrectomy handpiece
is connected but not
initialized.
Follow the instructions
on the screen to
initialize and verify
the operation of the
handpiece
314
Please, prime (of the
I/A cassette)
The I/A cassette is
connected but not
initialized.
Follow the on-screen
instructions to
initialize and verify
the operation of the
I/A cassette
318
Select peristaltic to
aspirate
In "Fluid" mode, while
injecting silicone, the
Venturi pump can not be
used.
Select the peristaltic
pump to aspirate
while injecting
silicone oil.
400 Other Messages
Id
Message
Possible cause
Correction
measurement
400
"Do you want to skip
the priming?"
One of the pedal switches
has been pressed, the
phaco handpiece is
connected but not
initialized.
a) The user
inadvertently
activates the foot
switch. Remove the
foot and press "NO" on
the screen
b) Faulty footswitch.
Replace the pedal