Chapter 5 - About capturing an image
Note
l
Eye steering is not available for every procedure.
l
Full user details can be found in the User Guide which is installed with the V2 Vantage Pro
applications.
View captured image
The image is identified by the eye and timestamp details. You can:
l
Check the quality of the full sized image.
l
Central pole shows the close-up image of the optic disk and macula. Once you are in this close-up
mode you can drag the image to review other areas in this close-up view. Click, and hold down, the
left mouse button and drag the image to the area you want to view. The Central pole button changes to
the Full field button when pressed. Click Full field to show the full image.
l
Scroll through the series of images by clicking Previous image or Next image. Check the quality of each
image.
l
Discard any images that are not required by clicking Discard. Images will be discarded at the end of
the session.
Click View Image Series to display all the thumbnail images. You can end the session from the Image Series view.
Checking the quality of captured images
You should check the quality of each captured image to determine if the exposure settings need to be
changed to suit the patient’s individual characteristics.
Captured image thumbnails are tiled and displayed at the side of the main image display area. Click an
image thumbnail to display it in the main image viewing area.
l
Poor quality image example, see Badly aligned images - examples on the facing page.
Lids, lashes obscuring the image
You may find that lids and lashes cover part of the image, obscuring some areas of the retina. This may
happen if the patient was too far away from the device, or blinked, when the image was captured. Use the
External Eye Image to check the patient alignment, see Checking the Exterior Eye Image.
Lids and lashes may sometimes obscure part of the image. You can reduce this by following your practice
procedures, see Minimizing lids and lashes in images on page 33.
Patient positioning
You can improve the quality of a captured image by properly positioning the patient. There are simple
ways to improve the patient's positioning.
Capturing an image from a well-positioned patient should produce a good quality image, see Good
Quality Image example below.
Note
Sometimes the patient may be misaligned, see Badly aligned images - examples on the facing page.
Good Quality Image example
The following is an example of an optomap® image of a well-positioned eye. This is the type of image you
should be trying to achieve.
If you do not capture an image like this you should check the instructions for capturing an image, see
Capturing images on page 21.
Note the following:
30 of 100