SERVICE MANUAL
Set-up
Ow-RX (Rev. 1)
68
The following table displays the values of programmable
parameters
Display
Parameters Set Value New setting WARNING
P00
Stand-by timeout 2 min
P01
Duty cycle 60
(1)
P02
Exposure Counter ////
P03
Film sensitivity 1.00
P04
Long cone OFF
P05
Display back light ////
P06
Dose Correction Factor ////
P07
Language ////
P08
Manufacturer's Logo Show logo
P09
Disable X-Rays
X-ray enabled
(ON)
P10
Display mode reverse OFF
P11
LCD Contrast Adjustment ////
P12
Beep on keypress ON
P13
Exposure extra time 0
(1)
P14
Custom ID ///
P15
Tube type Toshiba
P16
Restore Factory Settings ////
WARNING:
Parameters shown with (1) in the previous table can not be changed; the
perfect running of the system or its compliance with Directive CE 93/42
cannot be guaranteed if any change is made.