5.Menu Interpretation
45
Use the current marker readings to set the parameters (such as
Center frequency, Reference level).
5.6.3.1 [MkrCF]
Sets the center frequency of the analyzer to the frequency of the
current marker. This feature quickly moves the signal to the center
of the screen.
In Normal marker mode, the center frequency will be set to the
frequency of the current marker.
In Delta marker mode, the center frequency will be set to the
frequency at which the Delta Marker is located.
The function is not available in zero span mode.
5.6.3.2 [MkrCF Step]
Sets the center frequency step of the analyzer to the frequency of
the current marker.
In Normal marker mode, the center frequency step will be set to
the frequency of current marker.
In Delta marker mode, the center frequency step will be set to
the frequency at which the Delta Marker is located.
The function is not available in zero span mode.
5.6.3.3 [MkrStart]
Sets the start frequency of the analyzer to the frequency of the
current marker.
In Normal marker mode, the start frequency will be set to the
frequency of the current marker.
In Delta marker mode, the start frequency will be set to the
frequency at which the Delta Marker is located.
The function is not available in zero span mode.
5.6.3.4 [MkrStop]
Sets the stop frequency of the analyzer to the frequency of the
current marker.
In Normal marker mode, the stop frequency will be set to the
frequency of the current marker.
In Delta marker mode, the stop frequency will be set to the