Use the current marker readings to set the parameters (such as
Center frequency, Reference level).
5.6.3.1 [Mkr→CF]
Sets the center frequency of the analyzer to the frequency of the
current marker. This feature quickly moves the signal to the center
of the screen.
⚫ In Normal marker mode, the center frequency will be set to the
frequency of the current marker.
⚫ In Delta marker mode, the center frequency will be set to the
frequency at which the Delta Marker is located.
⚫
The function is not available in zero span mode.
5.6.3.2 [Mkr→CF Step]
Sets the center frequency step of the analyzer to the frequency of
the current marker.
⚫ In Normal marker mode, the center frequency step will be set to
the frequency of current marker.
⚫ In Delta marker mode, the center frequency step will be set to
the frequency at which the Delta Marker is located.
⚫ The function is not available in zero span mode.
5.6.3.3 [Mkr→Start]
Sets the start frequency of the analyzer to the frequency of the
current marker.
⚫ In Normal marker mode, the start frequency will be set to the
frequency of the current marker.
⚫ In Delta marker mode, the start frequency will be set to the
frequency at which the Delta Marker is located.
⚫ The function is not available in zero span mode.
5.6.3.4 [Mkr→Stop]
Sets the stop frequency of the analyzer to the frequency of the
current marker.
⚫ In Normal marker mode, the stop frequency will be set to the
frequency of the current marker.
⚫ In Delta marker mode, the stop frequency will be set to the