5.14 【System】 hardkey
Key access:
Press Shift key, then System hardkey (2 in numeric keypad).
The menu for system parameter settings includes [System >],
[Setting >], [PowerOn/Preset >], [Save/Recall >]. For first time you
use the spectrum analyzer, set the system settings, the system will
store the settings, restart the machine after power off won’t change
the settings.
5.14.1 [System >]
Access system information, firmware update, and option submenu.
5.14.1.1 [System Info]
Display the current system information, such as serial number,
firmware version, temperature, MAC address, etc.
5.14.1.2 [Firmware Update]
To update your instrument firmware, do the following:
1. Create a folder named “spectrum” (lowercase) on the root
directory of the USB memory device, and copy the firmware file
onto this folder.
2. Insert the USB memory device into the top panel USB connector
on your instrument. Press Shift key, then System hardkey.
Press [System >], and press [Firmware Update] to execute
firmware update.
3. The analyzer will perform the update process. The upgrade
procedure will take approximately 30 seconds. During the
update process, do not remove the USB memory device, do not
power off the instrument or press any key. If the update process
fails, please report the problem to your distributor or our
technical support.
4.Once the upgrade is completed, the instrument will automatically
restart.
5.14.1.3 [Option >]