Operator's Manual
Issue 14 / Mar 2016
/ UMC0071
MercuryiPS
©2016 Oxford Instruments NanoScience. All rights reserved.
Page 119
MAN SERL R
Reads the serial number of the daughter
card
NICK R/W Sets the name of the device
STAT R
Reads alarm flags for the device (Hex
Format)
TYPE DUM:CUR:VOLT:RAT R/W Sets/Reads the pressure transducer type
EXCT TYPE None:10V:15V R/W
Sets/Reads the pressure transducer
excitation
CAL OFFS 0 - 1000 R/W Sets the offset of the calibration curve
CAL SCAL 0 - 1000 R/W Sets the scale of the calibration curve
CAL FILE R/W
Sets the calibration file to use to calculate the
temperature
CAL INT NONE:LIN R/W
Sets the interpolation method for the
calibration file
CAL HIGHL 0 - 1000 R/W
Set the maximum value for temperature
setpoint (hot limit)
CAL LOWL 0 - 1000 R/W
Set the minimum value for temperature
setpoint (cold limit)
CAL CAL W Calibrates the hardware
The signals for a pressure board are DEV:<UID>:PRES: followed by:
COMMAND OPTIONS Read/Write DESCRIPTION
SIG VOLT R Sensor voltage
SIG CURR R Sensor current
SIG PRES R Measured pressure
SIG POWR R Sensor power dissipation
10.3.5.7 Addressing a pressure control loop
To address a control loop, you must specify the pressure sensor that is associated with the
control loop:
DEV:<UID>:PRES:
where <UID> is a unique identifier that is allocated to each board, based on its SPI location.
The configuration settings for a control loop are DEV:<UID>:PRES: followed by:
COMMAND OPTIONS Read/Write DESCRIPTION
LOOP AUX UID R/W
Assign Auxiliary device to
Temperature
LOOP SWFL R/W
Sets the file to read from for the
Sweep Table
LOOP SWMD OFF:ON R/W Sets the sweep mode
LOOP TSET 0 - 2000 R/W Sets the pressure set point
LOOP FSET 0 - 100 R/W
Sets the flow percentage (Manual
flow)
LOOP FAUT OFF:ON R/W Enables/Disables flow control