EasyManua.ls Logo

Oxford Instruments NanoScience MercuryiPS - Pressure Board Functionality; Expansion Slot Layout and Positions

Oxford Instruments NanoScience MercuryiPS
169 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator's Manual
Issue 14 / Mar 2016
/ UMC0071
MercuryiPS
©2016 Oxford Instruments NanoScience. All rights reserved.
Page 22
Pressure
This is a board which measures signals from a pressure transducer. It can be used to control
pressure in a cryogenic system. A maximum of eight of these boards can be fitted. Chapter
13 gives a full technical description.
The iPS automatically identifies every daughter board present.
Figure 4 Layout of MercuryiPS expansion slots
The location of daughter boards is subject to certain constraints, as given in the following table.
Board Allowed position Typical position
Level sensor Any 5
Auxiliary I/O Any 4
Temperature sensor Any 6, 7 and/or 8
GPIB GPIB only
Pressure Any 8

Table of Contents