Appendix
pco.edge User Manual V1.03 © PCO AG, Germany
61
Operation
First connect power out of cooling unit with power in of camera with the
PCO WAT camera cable.
The cooling unit provides two operation modes. Please select.
Operation Mode “on”: the cooling unit is on permanently and
provides power to the camera. Camera can be switch on and
off as necessary.
Operation mode “follow”: the cooling unit turns on when the
camera is switched on and vice versa.
Error Codes
The cooling liquid temperature sensor is located in the reservoir (tank).
27°C fan turns off
36°C fan turns on
55°C warning message
60°C error message
If a warning level is passed, the Power LED blinks slowly and the Error LED
displays the error code. If a failure level is passed, the Power LED blinks
fast and the Error LED shows the error code.
Note: the camera has its own protection circuit and will shut down
automatically when the electronics temperature exceeds safety level.