A200SP Absorber 22 User Manual
6. Pre-use Checks
6.1 Operators pre-use checklist
WARNIN
Before you test and ths equpment, make sure that you
have read ths user manual
If the system fals a test, do not use the equpment The
Absorber must be checked by a traned engneer
omplete all tests pror to frst daly use, complete tests 5 to
12 before each clncal procedure
Test Reference Remarks
1 Inspect hoses, seals and power cables for
damage
NOTE
1 The absorber must be attached to an anaesthetc machne, whch must be leak tested before the checks are carred out
2 The procedures detaled n sectons 62 to 66 must be carred out n the order lsted
3 heck that the manometer s zeroed before use
2 Vsual nspecton of the absorbant
Inspect the absorbant, replace f
necessary
3 Absorber securty check secton 534
Before refttng the canster,
check that the sealng surfaces
are clean and dust free
Ensure that the canster seals
securely when reftted
4 Breathng crcut hose nstallaton check secton 522
heck that the fresh gas hose
s connected to the anaesthetc
machne
5 Absorber leak test secton 62
6 APL valve functon check secton 63
7 NRV functon check secton 64
8 Bag/Vent swtch test secton 65 heck for correct operaton
9 Heater test secton 55
10 Leak test wth canster removed secton 66
11 Absorber leak test secton 62
12 arry out all Anaesthetc workstaton pre-use
checks
Refer to applcable workstaton IFU