A200SP Absorber 33 User Manual
Sterilisation
8.5 Sterilisation and Disinfectant Treatment
Table
NOTE
1 After chemcal dsnfecton, always rnse wth warm water
and dry n ar
2 Do NOT clean any component n an automatc cleanng/
washng machne
3 Before clncal use, ALWAYS carry out the Pre-use hecks
lsted n secton 6 of ths manual
4 Refer to the nstructons for use for the autoclave
equpment
8.5.1 Absorber
omponent Soap
water
dex sonacd
(see NOTE 1
above)
Steam
autoclave
Maxmum temperature
°F °
Breathng hoses (check
manufacturer’s nstructons)
Yes Yes Yes 278 137
Safelock fttngs Yes Yes Yes 278 137
Reservor bag (check manufacturer’s
nstructons)
Yes Yes Yes 278 137
Manfold block (ncludng non-return
valves) and frame assembly
Yes Yes Yes 255 124
*Absorber canster Yes Yes Yes 255 124
*Absorbent contaner Yes Yes Yes 255 124
*arrer (for pre-pack) Yes Yes Yes 278 137
APL valve Yes Yes Yes 278 137
Pressure gauge No No No - -
PEEP valve Yes Yes Yes 278 137
Electrcal nterface unt No No No - -
Heater (opton) No No No - -
*Reft the absorbent canster (wth nner carrer and contaner,
f used) to the absorber assembly before autoclavng
NOTE
Durng autoclavng, slght dscolouraton of the manfold block,
absorbent canster, and absorbent contaner may occur
Ths s not an ndcaton of component deteroraton
8.5.2 Ventilator Bellows
Item Method
Bellows as, lqud, autoclave (20 cycles maxmum)
Hoses as, lqud, autoclave
O-rngs as, lqud, autoclave
Bellows base as, lqud, autoclave
Exhalaton valve as, lqud, pasteurse, low temperature autoclave
Bellows canster Lqud, autoclave