A200SP Absorber 38 User Manual
Appendix
Guidance and manufacturer’s declaration - electromagnetic immunity
The A200SP Absorber s ntended for use n the electromagnetc envronment specfed below The customer or user of the
A200SP Absorber should assure that t s used n such an envronment
Requred test 60601 test levels for equpment used
n a professonal healthcare faclty
envronment only
omplance level
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2, 4, 8, 15 kV air
± 8 kV contact
± 2, 4, 8, 15 kV air
Radiated RF EM Fields
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
Proximity fields from RF
wireless communications
IEC 61000-4-3
Frequency - Test level
385 MHz - 27 V/m
450 MHz - 28 V/m
710 MHz – 9 V/m
745 MHz – 9 V/m
780 MHz – 9 V/m
810 MHz - 28 V/m
870 MHz - 28 V/m
930 MHz - 28 V/m
172 Hz - 28 V/m
1845 Hz - 28 V/m
197 Hz - 28 V/m
245 Hz - 28 V/m
524 Hz – 9 V/m
550 Hz – 9 V/m
5875 Hz – 9 V/m
Frequency - Test level
385 MHz - 27 V/m
450 MHz - 28 V/m
710 MHz – 9 V/m
745 MHz – 9 V/m
780 MHz – 9 V/m
810 MHz - 28 V/m
870 MHz - 28 V/m
930 MHz - 28 V/m
172 Hz - 28 V/m
1845 Hz - 28 V/m
197 Hz - 28 V/m
245 Hz - 28 V/m
524 Hz – 9 V/m
550 Hz – 9 V/m
5875 Hz – 9 V/m
Electrcal fast transent /
burst
IE 61000-4-4
± 2 kV for power supply lnes
± 1 kV for nput / output lnes
± 2 kV for power supply lnes
Surge
IE 61000-4-5
± 05, 1, 2 kV lne(s) to earth
± 05, 1 kV lne(s) to lne(s) for power supply
lnes
± 2 kV lne(s) to earth for nput / output lnes
± 05, 1, 2 kV lne(s) to earth,
± 05, 1 kV lne(s) to lne(s) for power supply
lnes
onducted RF
IE 61000-4-6
3 V - 150 kHz to 80 MHz
6 V - ISM bands between 150 kHz to 80 MHz
3 V - 150 kHz to 80 MHz
6 V - ISM bands between 150 kHz to 80 MHz
Voltage dips and interruptions
IEC 61000-4-11
0% UT (100% dp n UT) for 05 cycle
0% UT (100% dp n UT) for 1 cycle
70% UT (30% dp n UT) for 25/30 cycles
0% UT (100% dp n UT) for 250/300 cycles
0% UT (100% dp n UT) for 05 cycle
0% UT (100% dp n UT) For 1 cycle
70% UT (30% dp n UT) for 25/30 cycles
0% UT (100% dp n UT) for 250/300 cycles
Power frequency (50 Hz)
Magnetic field
IEC 61000-4-8
30 A/m 30 A/m