Instrument Performance Validation . 103
7. Make sure that the Validation Details are correct, and add any missing information.
NOTE: The serial number of your instrument is usually printed on a silver label in the
lower sample compartment area. The serial number of the TRM is on the TRM
sample holder.
8. Click OK.
The information that will be printed on your Certificate of Performance Validation is
displayed.
NOTE: There are two peaks, with details given in both nm and cm
-1
.
9. Make sure that all the details are correct.
Peaks found outside the
tolerance of the nominal value are displayed in red. If a
peak is not found within 10 cm
-1
of the nominal value, No Peak is displayed in red
(see
Possible Reasons for Validation Failure
on page 104).
10. To save the information on the certificate, click Save.
The Specify IPV Result File dialog is displayed.
11. Choose the Drive, Folder and File Name.
12. Click OK.
The file is saved as an .ipv file. The Specify IPV Result File dialog closes and the
certificate information is redisplayed.
To print the certificate, see page 104.