9.3 Power
Adjusts the microphone’s transmission power (aecting
its operating range) in order to extend the battery’s
operating time.
How to use it:
• Select Settings > Power
Select function required:
Standard (default) (100 mW, range: 20 m / 65 feet)
Reduced (16 mW, range: 12 m / 39 feet)
9.4 Reconguration
Resets Roger Touchscreen Mic’s language settings to
factory “out of the box” settings.
How to use it:
• Select Activate/Deactivate (default)
When activated, you will be asked to select the
microphone’s language the next time it is turned on.
When deactivated, the currently selected language will
be retained when microphone is turned o.
9.5 Verication mode
Used by professionals to verify a device according to the
Roger POP (Phonak Oset Protocol). When Verication
mode has been activated, an orange banner will be
displayed below the status bar.
How to use it:
• Select Activate/Deactivate (default)
Verication mode is automatically deactivated when
microphone is rebooted (turned o and back on).
9.6 Quick menu
Locks most icons (see below) in order to avoid unintended
use. Requires a passcode to be activated. All icons will be
locked except: Connect, Check, Disconnect, Settings (since
these four icons are essential for daily use of the
microphone). Physical buttons will not be locked.
Connect Disconnect SettingsCheck

Table of Contents

Other manuals for Phonak roger

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Phonak roger and is the answer not in the manual?

Phonak roger Specifications

General IconGeneral
BrandPhonak
Modelroger
CategorySwitch
LanguageEnglish

Related product manuals