EasyManua.ls Logo

Planer BT37 MarkII - Page 150

Planer BT37 MarkII
175 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
150
BT37 MarkII: Service Manual MA200606 1.0.0
Configuring BT37 toolbox for manual calibration
20 Configuring BT37 toolbox for manual calibration
Configuring the BT37 toolbox for manual calibration will be based on the assumption tha
t
a single temperature measurement device with a single probe and a flowmeter will be
used.
A sample of the configuration form is shown below (please note that the calibration
offsets have been entered as "0" (as BT37 toolbox does not use them for manual
calibration).
Enter your equipment serial numbers as these will be displayed on your calibration
record.
Enter "-1" as the PPL number, as this will instruct BT37 toolbox that a manual input will
be required.

Table of Contents