Chapter H - PARTS REPLACEMENT & REPAIR
H-6 Planmeca Proline CC panoramic x-ray
Technical Manual
3.6 Layer hold limits
a) Open the cover of the lover shelf
b) Unscrew two fixing screws
c) Remove the connector
d) After installing new PCB check the limits and adjust if necessary
3.7 Primary position detector
a) Remove the front cover of tube head
b) Unscrew the two fixing screws.
c) Remove the connector
d) After installing new PCB check the radiation beam with all slots and adjust if necessary.