Chapter A - GENERAL & TECHNICAL DATA
Planmeca Proline CC panoramic x-ray A-3
Technical Manual
3 TECHNICAL SPECIFICATIONS
X-ray tube Toshiba D-052SB
Focal spot size 0.5 x 0.5mm
Target angle 5°
Total filtration 2.5 mm Al
Generator Constant potential, microprocessor controlled,
operating frequency 80 kHz
Anode voltage 60-80 kV ± 2.5 kV
Anode current 4-12 mA ± 1.0 mA
Exposure time 2.5 - 18 s as indicated ± 10%
Cooling period Automatically controlled
Film size 15 x 30 cm (film-based x-ray)
Cassette Flat (film-based x-ray)
SID 480 mm
Magnification Constant 1.2
Line voltage 100, 117, 220, 230-240 V~ 50 or 60 Hz
Regulation Automatic ± 10%
Line current max. 8A at 230V~, 15A at 100V~
Electrical classification
Class I
Type B
Mechanical data
Weight 108 kg
Dimensions (HxDxW) 2200x930x890 mm, 86.6x36.6x35 inch
Color RAL 9002
Environmental requirements
Ambient temperature Operating +5°C to +40°C
Storage -10°C to +50°C (film-based x-ray)
0°C to +50°C (Dimax3 digital x-ray system)
Humidity 25% - 75%
Dimax3 digital x-ray system
CCD pixel size 33 µm
Image pixel size 130 µm normal resolution
99 µm enhanced resolution
Image size 2200x1000 pixels normal resolution
2940x1360 pixels enhanced resolution
Sensor height Pan 9 x 136 mm
Ceph 9 x 230 mm
Manufacturer
PLANMECA Oy, Asentajankatu 6, 00810 Helsinki, FINLAND
phone: +358-0-7590 5500