Chapter D - ADJUSTMENT & CALIBRATION
D-2 Planmeca Proline CC panoramic x-ray
Technical Manual
1 REQUIRED TOOLS
Panoramic mode calibration
Alignment ruler (part number 50973). For checking the position of the patient positioning
mechanism.
Ball phantom (part number 50971). For checking the position of the patient positioning mech-
anism and the positioning lights.
Frankfort plane alignment tool (part number 50977). Used with the ball phantom for checking
the position of the Frankfort plane light.
Beam alignment tool (part number 50972). For checking the position of the panoramic x-ray
beam.
Alignrul.eps
Alignment ruler
START
EN
D
PLANMECA
No.50971
X-LINE
Y-LINE
Ball phantom
START
EN
D
PLANM
ECA
N
o.50971
X-LIN
E
Y-LINE
PLANMECA OY
NO.50977
Frankfort plane
alignment tool
Beam alignment tool