35
Chapter
E Calibrating the Sensor
1 ProOne setup
NOTE: Make sure that the third person “average adult” is selected and re-attach all c-arm
c
overs.
Press the kV/mA at the main screen. (Figure 53)
Figure 53
Set the unit to 64kV and 5mA. (Figure 54)
Figure 54
Printed copies of this document are considered uncontrolled.
7535.11.Rev001 08.07.2018