WMA-5 Operation Manual V. 1.01 27 support@ppsystems.com
Reset Zero Absorber
Reset Zero Absorber
Yes Button
Reset the Zero Absorber capacity gauge to 100%. This should be set every time
the soda lime is changed in the Auto-Zero column to ensure reporting accuracy.
No Button
Returns to the Settings 1 Menu without resetting the Zero Absorber capacity
gauge.
Flow Settings
This function allows the user to change/view the Flow rate of the WMA-5.
Flow Settings
Flow (cc/min)
The measured flow rate
Pump Power (%) Button
The Pump Power setting (0-100%)
Back Button
Returns to the Settings 1 Menu.