EasyManua.ls Logo

PP Systems WMA-5 - Measure Format Settings Table

Default Icon
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WMA-5 Operation Manual V. 1.01 63 support@ppsystems.com
Measure Format Settings Table
This table describes the six different Measure Formats. Note: In the examples, there CRC parameter is
turned off and there are no CRCs concatenated to the end of a message. <CR> represents a carriage
return (ASCII 13).
Measure Format Setting
Setting
Description
M1
M1 is the legacy data format used in the WMA-4 (previous model to the WMA-5).
Note: this format is NOT comma delimited.
An example is:
M1100012801131900968115025000000000000000000000099700<CR>
The format is as follows:
M1: Determines what data format is being displayed. In this case, M1 is WMA-4
format. (position 0,1)
WMA ID: 0-9. (Identifier for WMA-5 when multiple units are used). (position 2)
Record #: 1-9999. (position 3-6)
Day: 1-31. (position 7,8)
Month: 1-12 (January - December). (position 9,10)
Hour: 1-24 (based on 24 hour clock). (position 11,12)
Minutes: 0-59. (position 13,14)
CO2 Ref: CO
2
reading in parts per million (ppm = µmol m
2
s
-1
). (position 15-19)
mb Ref: H
2
O reading in millibars (mb). (position 20-22)
AnT: Temperature ( °C) of the gas sample. (position 23-26)
Input 1: millivolts (mV) signal from sensor connected to AUX1. (position 27-30)
Input 2: millivolts (mV) signal from sensor connected to AUX2. (position 31-34)
Input 3: not used. Value is 0000. (position 35-38)
Input 4: not used. Value is 0000. (position 39-42)
Input 5: not used. Value is 0000. (position 43-46)
ATMP: Atmospheric pressure in millibars (mb). (position 47-50)
ECode: Displays error code message. See Error Messages for the WMA-5 table
below for details. (position 51,52)
M2
M2 is the No Header format contains everything but the header and is comma
delimited.
An example is:
M2, 1094, 0.0, 20.41, 327, 0.0000, 55.0, 1004.2,0x0000<CR>
The format is as follows:
M2: Determines what data format is being displayed. In this case M2 is No Header
format.
CO
2
(ppm): CO
2
reading in parts per million (ppm = µmol m
2
s
-1
)
H
2
0 (mb): H
2
O reading
O
2
(%): O
2
reading
Flow Rate (cc/min): Rate of flow.
Aux Voltage (volts): AUX1.
H
2
0 Sensor Temperature (°C): Temperature of the gas sample.
Air Pressure (mb): Atmospheric pressure
System Error (hex): Displays error code message. See Error Messages for the
WMA-5 table below for details.
M3
M3 is the full format that has the complete data set and is comma delimited.

Table of Contents

Related product manuals