What to do if PRECISION DIGITAL PD6000-6R0 locks up?
- SStephanie HardinAug 17, 2025
If the PRECISION DIGITAL Measuring Instruments locks up or does not respond, try cycling the power to reboot the microprocessor.
What to do if PRECISION DIGITAL PD6000-6R0 locks up?
If the PRECISION DIGITAL Measuring Instruments locks up or does not respond, try cycling the power to reboot the microprocessor.
Why is the PRECISION DIGITAL Measuring Instruments display reading not accurate?
If the PRECISION DIGITAL Measuring Instruments display reading is not accurate, check the selected math function (Linear, square root, etc.) and the scaling or calibration.
How to fix unstable display on PRECISION DIGITAL Measuring Instruments?
If the PRECISION DIGITAL Measuring Instruments display is unstable, check the following: 1. Input signal stability and value 2. Display scaling vs. input signal 3. Filter and bypass values (increase)
Why PRECISION DIGITAL PD6000-6R0 relay doesn't respond to signal?
If the PRECISION DIGITAL Measuring Instruments relay and status LED do not respond to the signal, check the relay action in the Setup menu and the set and reset points.
What to do if PRECISION DIGITAL Measuring Instruments display reading fixed number?
If the PRECISION DIGITAL Measuring Instruments display does not respond to input changes and reads a fixed number, check the display assignment. It might be displaying max, min, or set point.
How to correct reversed relay operation in PRECISION DIGITAL Measuring Instruments?
If the PRECISION DIGITAL Measuring Instruments relay operation is reversed, check the fail-safe setting in the Setup menu and the wiring of the relay contacts.
Why PRECISION DIGITAL PD6000-6R0 display response is too slow?
If the PRECISION DIGITAL Measuring Instruments display response is too slow, check the filter and bypass values.
Why does PRECISION DIGITAL PD6000-6R0 Measuring Instruments display Hi and Lo?
If the PRECISION DIGITAL Measuring Instruments display alternates between Hi and a number or Lo and a number, press the Menu button to exit max/min display readings.
Why PRECISION DIGITAL PD6000-6R0 Measuring Instruments displays 1.999999?
If the PRECISION DIGITAL Measuring Instruments meter displays 1.999999 or -99999, check the input selected in the Setup menu and the corresponding signal at the Signal connector.
How to unlock PRECISION DIGITAL PD6000-6R0 Measuring Instruments?
If you are not able to change setup or programming on the PRECISION DIGITAL Measuring Instruments and 'Locd' is displayed, the meter is password-protected. Enter the correct six-digit password to unlock it.
Model | PD6000-6R0 |
---|---|
Display Type | LCD |
Digits | 6 |
Input Type | RTD |
RTD Excitation Current | 0.5 mA |
Approvals | CE |
Mounting | Panel Mount |
Power Supply | 85-265 VAC |
Operating Temperature | 0 to 50 °C |
Temperature Stability | ±0.01% of span/°C |
RTD Type | Pt100 |
Power | 5W |
Housing Material | Plastic |
Dimensions | 96 x 48 x 100 mm |