PD6000 Analog Input Process Meter
Instruction Manual
0-20 mA, 4-20 mA, 0-5 V, 1-5 V, and ±10 V Inputs
NEMA 4X, IP65 Front
Universal 85-265 VAC or 12/24 VDC Input Power
Large Dual-Line 6-Digit Display, 0.60" & 0.46"
Isolated 24 VDC @ 200 mA Transmitter Power Supply
Math Functions for Flow & Round Horizontal Tanks
Programmable Display & Function Keys
Sunlight Readable Display
32-Point, Square Root, or Exponential Linearization
Multi-Pump Alternation Control
2 or 4 Relays + Isolated 4-20 mA Output Options
External 4-Relay & Digital I/O Expansion Modules
RS-232, RS-422/485 Serial Communication Options
-40 to 65°C Operating Temperature Range
PRECISION DIGITAL CORPORATION
19 Strathmore Road • Natick MA 01760 USA
Tel (800) 343-1001 • Fax (508) 655-8990
www.predig.com

Table of Contents

Questions and Answers

  • E
    Emily MillerAug 17, 2025
    What to do if PRECISION DIGITAL PD6000-6R0 locks up?
    • S
      Stephanie HardinAug 17, 2025
      If the PRECISION DIGITAL Measuring Instruments locks up or does not respond, try cycling the power to reboot the microprocessor.
  • M
    myerspatrickAug 19, 2025
    Why is the PRECISION DIGITAL Measuring Instruments display reading not accurate?
    • T
      Travis BondAug 20, 2025
      If the PRECISION DIGITAL Measuring Instruments display reading is not accurate, check the selected math function (Linear, square root, etc.) and the scaling or calibration.
  • J
    Jessica WilliamsAug 22, 2025
    How to fix unstable display on PRECISION DIGITAL Measuring Instruments?
    • J
      Jean RichardsonAug 22, 2025
      If the PRECISION DIGITAL Measuring Instruments display is unstable, check the following: 1. Input signal stability and value 2. Display scaling vs. input signal 3. Filter and bypass values (increase)
  • H
    Heather CollinsAug 26, 2025
    Why PRECISION DIGITAL PD6000-6R0 relay doesn't respond to signal?
    • S
      Sophia MezaAug 26, 2025
      If the PRECISION DIGITAL Measuring Instruments relay and status LED do not respond to the signal, check the relay action in the Setup menu and the set and reset points.
  • A
    ajohnsonAug 28, 2025
    What to do if PRECISION DIGITAL Measuring Instruments display reading fixed number?
    • M
      Melissa SmithAug 28, 2025
      If the PRECISION DIGITAL Measuring Instruments display does not respond to input changes and reads a fixed number, check the display assignment. It might be displaying max, min, or set point.
  • T
    taylorchapmanAug 30, 2025
    How to correct reversed relay operation in PRECISION DIGITAL Measuring Instruments?
    • C
      Connor GarciaAug 30, 2025
      If the PRECISION DIGITAL Measuring Instruments relay operation is reversed, check the fail-safe setting in the Setup menu and the wiring of the relay contacts.
  • A
    Amy MackSep 2, 2025
    Why PRECISION DIGITAL PD6000-6R0 display response is too slow?
    • K
      keyamberSep 2, 2025
      If the PRECISION DIGITAL Measuring Instruments display response is too slow, check the filter and bypass values.
  • K
    Kelly KingSep 5, 2025
    Why does PRECISION DIGITAL PD6000-6R0 Measuring Instruments display Hi and Lo?
    • L
      ltaylorSep 5, 2025
      If the PRECISION DIGITAL Measuring Instruments display alternates between Hi and a number or Lo and a number, press the Menu button to exit max/min display readings.
  • M
    Mitchell HernandezSep 7, 2025
    Why PRECISION DIGITAL PD6000-6R0 Measuring Instruments displays 1.999999?
    • F
      Frank GonzalezSep 7, 2025
      If the PRECISION DIGITAL Measuring Instruments meter displays 1.999999 or -99999, check the input selected in the Setup menu and the corresponding signal at the Signal connector.
  • S
    Stephanie EvansSep 12, 2025
    How to unlock PRECISION DIGITAL PD6000-6R0 Measuring Instruments?
    • J
      John MoraSep 12, 2025
      If you are not able to change setup or programming on the PRECISION DIGITAL Measuring Instruments and 'Locd' is displayed, the meter is password-protected. Enter the correct six-digit password to unlock it.

PRECISION DIGITAL PD6000-6R0 Specifications

General IconGeneral
ModelPD6000-6R0
Display TypeLCD
Digits6
Input TypeRTD
RTD Excitation Current0.5 mA
ApprovalsCE
MountingPanel Mount
Power Supply85-265 VAC
Operating Temperature0 to 50 °C
Temperature Stability±0.01% of span/°C
RTD TypePt100
Power5W
Housing MaterialPlastic
Dimensions96 x 48 x 100 mm