3270 Series Operation Manual 61
BATT TIME: setting the Battery test time, the Left 5 digit monitor display the
“BATT” ,the right upper 5 digit monitor display the “TIME”, and right lower monitor
display setting value, the unit is "S". The range is 1S ~99999S.
BATT VTH: the Left 5 digit monitor display the “BATT” ,the right upper 5 digit
monitor display the “VTH”, and right lower monitor display setting value, the unit is
"V". The range is 0.01V to the full scale of the Voltage specification.
Press the Item key to enter the Item setting mode BATT PRESS START, the LED
indicator is ON, and then press the setting key. The LED indicator is ON. To exit the
setting, press the EXIT key and select LIN MODE. The setting sequence is as follows:
BATT PRESS START
BATT MODE LIN
BATT LIN
BATT TIME
BATT VTH
BATT LIN : setting the BATT LIN, the Left 5 digit monitor display the “BATT” ,the
right upper 5 digit monitor display the “LIN”, and right lower monitor display setting
value, the unit is "A". The range is 0.001A to the full scale of the CC mode
specification.