EasyManuals Logo

ProMinent delta User Manual

ProMinent delta
47 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
A container with water, which is free of additives that could falsify
the measured result, is needed for calibration. Immerse the sensor
removed that is still connected to the control module
delta
®
electri‐
cally into this water. Stir the sensor around the water for approx. 5
minutes until the reading on the on the control module delta
®
is
displayed steady and close to
„0“
.
1. Select the calibration menu
[Sensor Calibration]
. The press
the button
2. Select the unit
„Adj. zero point“
to be calibrated using the
buttons
or
3.
Continue with
ð
A prompt is shown in the display
4. Adapt the
„ Zero point“
value displayed during the calibration
using the buttons
or and when necessary, accept the
value using the button
ð
Refer to the Error Message table should an error be
displayed
NOTICE!
Then definitively calibrate the gradient with a suitable
reference method (e.g. DPD. titration etc.).
Calibrating the chlorine sensor: Zero
point
Measured variables for the control module delta®
39

Table of Contents

Other manuals for ProMinent delta

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent delta and is the answer not in the manual?

ProMinent delta Specifications

General IconGeneral
Stroke ratemax. 180 strokes/min
TypeDiaphragm metering pump
MaterialsPVDF, PTFE, stainless steel
Protection classIP 65
Diaphragm materialPTFE
DriveSolenoid
Power supply230 V, 50/60 Hz

Related product manuals