EasyManuals Logo

ProMinent delta User Manual

ProMinent delta
47 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
9 Technical data, maintenance, disposal
Voltage input
Value
Measuring range 0/4 mA - 25 mA (at 50 Ohm
measurement resistor)
Accuracy After calibrating ± 0.5% of the
measuring range transmitting
value at calibration temperature
Resolution 10-12 Bit
Input voltage protected against incorrect polarity and return of
electric energy to ± 30 V.
Switchable voltage output
Value
Output voltage 22.5V-26V load dependent, <
50 Ohm; maximum 50mA
Output voltage protected against incorrect polarity and return of
electric energy to ± 30 V.
Galvanic Isolation to the delta
®
front panel. Load current limiting to
approx. 55mA (51 mA - 58 mA).
Deactivation with short circuit (approx. 70 mA) due to foldback and
by software. Reactivation using software.
The control module delta
®
is maintenance free.
NOTICE!
Regulations governing disposal of used parts
Note the current national regulations and legal
standards which apply in your country
ProMinent Dosiertechnik, Heidelberg / Germany is prepared to
take back the decontaminated and cleaned used parts.
The current valid Declaration of Decontamination can be obtained
as download at
www.prominent.com
.
Electrical data
Maintenance
Disposal of Used Parts
Technical data, maintenance, disposal
46

Table of Contents

Other manuals for ProMinent delta

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ProMinent delta and is the answer not in the manual?

ProMinent delta Specifications

General IconGeneral
Stroke ratemax. 180 strokes/min
TypeDiaphragm metering pump
MaterialsPVDF, PTFE, stainless steel
Protection classIP 65
Diaphragm materialPTFE
DriveSolenoid
Power supply230 V, 50/60 Hz

Related product manuals