DFM 6.1 INSTRUCTION MANUAL
50
Register
Address
Description Register Type Format Type Comments
30904 Run Hours Input Register Floating Point Register (1 of 2)
30905 Run Hours Input Register Floating Point Register (2 of 2)
30923 Sensor Status Input Register Index (0-10) 0 = Sensor Good
4 = Sensor Open
5 = Sensor Short
7 = Low Signal
30925 Logging Status Input Register Index (0-2) 0 = Stopped
1 = Active
2 = Full
30926 Logging Used - % Input Register Floating Point Register (1 of 2)
30927 Logging Used - % Input Register Floating Point Register (2 of 2)
30962 Confidence Input Register Integer
30963 Direction Input Register Integer
30964 Pot 1 Input Register Integer
30965 Peak to Peak Input Register Integer
30966 Correlation Strength Input Register Integer