Pulsar Nova® sound level meter user manual 
 
Page 55 
2.  The battery, type 6F22 (PP3) can now be eased out of its holder and replaced.  
The  battery  should  be  eased  out  terminal  side  first  by  pushing  against  the 
spring at the other end.   
 
Ensure that the battery is inserted with the correct polarity with the negative terminal at 
the contact with the larger cut out. 
 
Battery type 
 
The battery should be an alkaline battery, not an ordinary dry cell.  The battery is 9 volts 
when new and will operate the calibrator down to 6.4 volts.   
 
When the battery voltage is below 6.6 volts but above 6.4 volts, the power LED will flash 
to indicate that the battery voltage is low. When the battery voltage is below 6.4 volts 
the calibrator will not turn on.   
 
A discharged battery may allow switch-on but will soon drop in voltage and indicate low 
battery or switch off. 
Specifications 
 
Frequency          1kHz ± 1% 
Sound Level          94dB  re 20μPa 
Standardisation        Model 105 - IEC 60942:2003 Class 1 
            Model 106 - IEC 60942:2003 Class 2 
Distortion          Less than 2%  
Operating Humidity        25 to 90% Relative Humidity 
Operating Static Pressure      65kPa to 108kPa 
Operating Temperature      -10
o
C  to +50
o
C 
Storing Temperature       -20
o
C  to +60
o
C 
Effective Volume        6.19 cm3 ± 0.2 cm
3 
Cavity Diameter        0.525 inch 
Battery          1 x 9v 6F22 (Neda 1604) 
Battery Life          Approx 15 Hours Continuous Use 
Battery Voltage        9v Nominal (10v Max, 6.4v Min) 
Weight With Battery        185g 
Dimensions          135mm x Ø48mm 
 
Technical Information 
 
The normal mode of operation of the calibrator is with the unit switched on. 
When  the  LED  indicates  the  unit  is  switched  on  this  produces  the  greatest  radio 
frequency emissions. 
 
The calibrator continues to function after exposure to contact discharges up to 4kV and 
air  discharges  up  to  8kV,  for  both  positive  and  negative  voltages  relative  to  earth 
ground. The calibrator conforms  to  IEC 60942:2003  for a modulated  root-mean-square 
electromagnetic field strength of 10 V/m. 
 
The maximum susceptibility to power and radio frequency fields is with the cavity facing 
away  from  the  emitter  with  the  battery  compartment  facing  the  table,  the  antenna 
polarisation horizontal and the calibrator switched on.