Pulsar Nova® sound level meter user manual 
 
Page 53 
 
This  chapter  refers  to  the  use  of  a  Pulsar  Instruments  plc  Model  105  or  Model  106 
Acoustic Calibrator. 
 
Switching on the Calibrator 
 
Press the Power Button on the end of the 
Calibrator  to  switch  the  unit  on.  The 
Indicator  will  illuminate  to  show  that  the 
unit  is  operating.  The  calibrator  will 
automatically switch off after 5 minutes to 
preserve battery power. 
 
To  switch  off  the  calibrator  manually, 
press  the  power  button  again  and  the 
indicator  will  extinguish  to  show  that  the 
unit is switched off. 
 
Permanent ‘On’ Mode 
 
For  some  applications  there  may  be  a  need  to  have  the  calibrator  switched  on 
continuously. To allow for this, the calibrator can be turned on by pressing and holding 
the  power  button  for  three  seconds.  Release  the  button  and  the  indicator  will  flash  to 
show that the unit is in permanent ‘ON’ mode. Press the power button to switch off the 
calibrator. 
 
Calibrating a Sound Level Meter 
 
Push the microphone of the sound level meter into the cavity at the end of the calibrator. 
Ensure the microphone is fully inserted into the cavity and is past the ‘O’ ring seals.  
 
The  microphone  should  be  parallel  to  the  body  of  the  calibrator.    Also  ensure  that  the 
small  bleed-hole  next  to  the  microphone  cavity  is  not  blocked  as  this  could  cause 
damage to the microphone.