EasyManua.ls Logo

Pulsar NOVA - Free Field Correction

Pulsar NOVA
61 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pulsar Nova® sound level meter user manual
Page 56
Free Field Correction
When calibrating a microphone which is to be used for free field measurements, a small
correction may be necessary to compensate for the difference between the microphone’s
free field response at ‘zero degrees’ or ‘head-on’ incidence and the pressure level
generated by the calibrator.
The correction is typically -0.3dB for ½ inch microphones (making the effective
calibration level 93.7dB).
Calibration corrections are listed below for the Pulsar Instruments plc ½” Capsules and
three microphone capsules commonly used in Calibration Laboratories:
Microphone Correction Values
Microphone Type Calibration Correction Effective Calibration Level
PM1 -0.3dB 93.7dB
PM2 -0.3dB 93.7dB
B&K 4134 0dB 94.0dB
B&K 4180 0dB 94.0dB
B&K 4192 0dB 94.0dB
Example:
An example of the procedure used to calculate the value for an PM1 microphone is
shown below :
Level = 94.0dB + Microphone Correction
Level = 94.0dB + ( -0.3dB)
Level = 93.7dB

Table of Contents

Related product manuals