Pulsar Nova® sound level meter user manual 
 
Page 56 
 
Free Field Correction 
 
When calibrating a microphone which is to be used for free field measurements, a small 
correction may be necessary to compensate for the difference between the microphone’s 
free  field  response  at  ‘zero  degrees’  or  ‘head-on’  incidence  and  the  pressure  level 
generated by the calibrator.  
 
The  correction  is  typically  -0.3dB  for  ½  inch  microphones  (making  the  effective 
calibration level 93.7dB).  
 
Calibration corrections  are listed  below for  the  Pulsar  Instruments  plc  ½”  Capsules and 
three microphone capsules commonly used in Calibration Laboratories: 
 
Microphone Correction Values 
 
Microphone Type     Calibration Correction  Effective Calibration Level 
PM1        -0.3dB       93.7dB 
PM2        -0.3dB       93.7dB 
B&K 4134      0dB        94.0dB 
B&K 4180      0dB        94.0dB  
B&K 4192      0dB        94.0dB 
 
 
Example: 
 
An  example  of  the  procedure  used  to  calculate  the  value  for  an  PM1  microphone  is 
shown below : 
 
Level = 94.0dB + Microphone Correction 
Level = 94.0dB + ( -0.3dB)  
Level = 93.7dB