GEM5000 Gas Analyzer
Page 27 of 102
2) Select ‘Key 5 – Adjust Flow Fail’ and the following screen is displayed:
3) Manual adjustment of the flow fail is available via this option and can be carried out
with use of ‘Key 4 – Scroll left’ () less sensitive and ‘Key 6 – Scroll right’ (>) more
sensitive.
4) Select the soft-key Save to store the setting or select soft-key Exit’ to exit the screen
without saving the change.
5) The operator will return to the ‘Operation settings’ menu.
Note: The default setting displays the bar in the centre. BEFORE altering this
setting, please contact Technical Support at (800) 968-2026 or email
landtec_support@qedenv.com
7.2.6 Technician login
This option enables the operator to select or change a pre-defined
technician login and all subsequent readings will be tagged with this
Technician Login ID. The technician ID must already have been created
using the LSGAM software and uploaded to the instrument.
1) Select the ‘Menu’ button on the front of the analyzer to display the ‘Device Settings’
menu. Press the soft key to display ‘User Settings’.
2) Select ‘Key 6 – Technician login’ and the following screen is displayed:
3) Use the cursor keys to move throughout the list of IDs displayed; select the ‘Enter’
key to select choice of ID, default if no IDs are listed or soft-key ‘Skip’ to skip the
selection.