GEM5000 Gas Analyzer
Page 60 of 102
9.6 Gas analyzer
For the GEM5000 gas analyzer the calibration options can be found by selecting the ‘Menu’
key followed by soft-key ‘Operation Settings’. Select ‘Key 2 – Gas Check’ then follow the
instructions on the analyzer screen by selecting ‘Key 2 – Gas Check’.
9.7 Calibration processes best practice
The following process diagrams outline the calibration steps.
Ensure that you are regulating calibration gas down to below 200 mbar pressure, if
you’re not using a Landtec regulator. The use of a pressure relief valve is also highly
recommended.
When calibrating, it is recommended to use a calibration mixture close to the levels
you are trying to measure, i.e. if you are trying to measure gas migration on a closed
landfill, we’d recommend calibrating with CH4 5%/CO2 5%.
In regards to frequency, we would recommend that you perform a fresh air calibration
before each monitoring session, and a span calibration typical every 4 6 weeks.