GEM5000 Gas Analyzer
Page 8 of 102
from being adversely affected, thus ensuring that the type of protection is not
compromised.
The relative pressure range is +/-500 mbar. Note, however, that the input pressure
should not exceed +/- 500 mbar relative to atmospheric pressure and the output
pressure should not exceed +/- 100 mbar relative to atmospheric pressure.
For CSA (Canada) the 5000 series of gas analysers are certified to Hazardous
Area Classification
CLASS 2258 03 - PROCESS CONTROL EQUIPMENT - Intrinsically Safe and Non-
Incendive Systems - For Hazardous Locations
Ex ib IIA:
Model GA 5000, GEM 5000 and BIOGAS 5000 Methane Detectors; portable, battery
powered with non-field-replaceable Battery Pack P/N 20087 or 2011113; intrinsically safe
and providing intrinsically safe circuits (“[ib]” for Zone 1) to Model GF5.2 Temperature
Probe (Connector B) and with entity output parameters as tabulated below; Temperature
Code T1; -10 ºC Tamb. +50ºC.
Note: This device has been investigated for electrical safety features only.
For CSA (USA) the 5000 series of gas analysers are certified to Hazardous Area
Classification
CLASS 2258 83 - PROCESS CONTROL EQUIPMENT - Intrinsically Safe and Non-
Incendive Systems – For Hazardous Locations - CERTIFIED TO U.S. STANDARDS
AEx ib IIA:
Model GA 5000, GEM 5000 and BIOGAS 5000 Methane Detectors; portable, battery
powered with non-field-replaceable Battery Pack P/N 20087 or 2011113; intrinsically safe
and providing intrinsically safe circuits (“[ib]” for Zone 1) to Model GF5.2 Temperature
Probe (Connector B) and with entity output parameters as tabulated below; Temperature
Code T1; -10 ºC Tamb. +50ºC.
Note: This device has been investigated for electrical safety features only.