16 65000159 Rev. 008 (06/17)
Calibration Procedure
Action
Instrument
Information
Illustration
When calibration is
complete, analyzer will
beep.
The display will confirm
whether calibration passed
or failed, and display date
and time.
Cal X Pass
10/05/16 11:00
or
Cal X Fail
10/05/16 11:00
or
Cal Fail
10/05/16 11:00
Insert FP
in Printer
The result will remain
displayed for one minute
after the door is opened.
See Troubleshooting if
calibration fails.
Open the analyzer door.
Analyzer will start beeping
and display “Insert
FastPack
®
in Printer” until
result is printed.
Remove and inspect the
FastPack
®
IP. Check the
inside of the analyzer door
for fluid. If there is fluid on
the FastPack
®
IP or
analyzer, disregard the
result and properly dispose
of the pack. Implement the
Cleaning Procedure.
Ensure that all chambers
have burst. Fluid should
easily flow between all
chambers. Observe that
most of the fluid in the
FastPack
®
IP is yellow. If
there is no yellow fluid,
disregard the result and
rerun a FastPack
®
IP.