65000159 Rev. 008 (06/17) 59
Troubleshooting the FastPack
®
IP System
Display
Cause
Action
PMT Failure
PMT Failure; Light Leak.
Ensure that the analyzer is not in direct
sunlight. Contact System Support.
Shutter Failure
Shutter Failure
Contact System Support.
Pressure Failure
System has a severe pressure leak. Air
pump not functioning.
Contact System Support.
Remove FastPack
A FastPack
®
IP is in the analyzer when
attempting to run diagnostics.
Remove FastPack
®
IP. Upon closing door,
diagnostics will automatically begin.
Temperature
Out of Range
Temperature too low may occur if door
is left open for long duration.
Close door. Recheck after 5 minutes.
If problem persists, contact System
Support.
Temperature too high may occur if
room temperature is too warm (>32°C
or 90°F).
Move analyzer to a cooler location or
wait approximately 15 minutes
between runs.
If problem persists, contact System
Support.
Temp Out of Range
Invalid Test
Temperature has exceeded correct
range during a test.
Discard pack. Retry test. If problem
persists, contact System Support.
Test Cancelled
Cancel was pressed during a test.
Cancelling a test will require the use of a
new FastPack
®
IP and new sample.
Thermistor Failure
Thermistor has failed.
Contact System Support.
Time Error
A time that is not possible has been
entered.
Re-enter correct time.