EasyManua.ls Logo

Quidel Solana - QC Samples; Manual Entry of Patient Samples; Barcode Entry of Patient Samples; Entry of Patient Samples When Using Virena

Quidel Solana
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46
QC Samples
If needed, select the appropriate tube.
Select QC from the Sample Type drop-down list.
An instruction to scan the QC barcode is displayed (Figure 43). Note: QC sample IDs are entered using a
barcode scanner only.
To enter the ID of the QC sample, touch the button to confirm the selection of a QC sample.
If QC sample type was entered in error, touch the exit button. The sample type reverts to Patient.
See instructions above for Patient Samples.
Note: QC samples identified as patient samples have result treated as patient samples, which
impacts Virena reporting, if applicable.
Scan the QC sample’s barcode.
If a barcode is scanned that is not a recognized ID for a QC sample, an error message is displayed.
Assign the ID to the QC tube.
Repeat as needed for additional QC samples.
Manual Entry of Patient Samples
If needed, select the appropriate tube.
If needed, select Patient from the Sample Type drop-down list.
If Virena is not activated touch the ID field and enter sample information.
Touch the Enter key. The ID will automatically populate the next available tube.
To enter additional sample information, repeat the steps above.
Touch the Play button to start the tests.
Touch the button to confirm all tubes are loaded and to start the run.
Barcode Entry of Patient Samples
If needed, select the appropriate tube.
If needed, select Patient from the Sample Type drop-down list.
Touch the ID field and scan sample barcode. The ID will automatically populate the next available tube.
To enter additional sample information, scan the next sample barcode. The ID will automatically populate
the next available tube.
Touch the Play button to start the tests.
Touch the button to confirm all tubes are loaded and to start the run.
Entry of Patient Samples When Using Virena
If Virena is activated, type the sample ID or scan the barcode (Figure 45).
Type Age (in years, using whole numbers).

Table of Contents