Optics
Feature
Specifications
Signal detection
Multiple wavelengths for up to 4 dyes per tube
Wavelength combinations
excitation/emission (nm)
470/520, 560/625, 520/570 and 625/680
Dyes
FAM™, ROX™, HEX™, CY®5, Coumarin, TAMRA™, etc.
Scan interval 12 tubes
10 seconds (for one wavelength channel)
Performance
1 raw data acquisition cycle for maximum 12 tubes with up to 4 colors (on 2
detectors) within 35 seconds
Sensitivity
10 to 12 mol/L fluorescein sodium in 0.1 N sodium hydroxide (measured
with excitation at 470 nm and detection at 520 nm)
Surveillance
Internal check among solid standards
Power Requirements of the Instrument
Feature
Specifications
External Power Supply
Input: 100-240V ~ 1.5-0.7A 47 63Hz
Output: 24V 2.5A
Installation/Overvoltage
Category
Power Supply Class I
Accessories
Feature
Specifications
External printer
Dymo 450 printer
Internal Barcode scanner
Maximum current 500 mA (USB 2.0 specification)
Control characters are CR/LF (carriage return line feed)