Set-up and Connection Procedures
Unpacking of Solana
Inspect the shipping container for obvious damage prior to opening. Unpack the contents, placing Solana
upright on a dry, stable, and level surface. Carefully examine the Solana for damage. Check that all other
system components are provided (see System Components section) and are in good condition.
In case of damage or any missing parts, contact Quidel Technical Support.
Placement of Solana
The Solana is portable and can be moved to a suitable location for testing or storage when locked properly.
When moving the instrument, lock the mechanical transport lock. See the Locking and Unlocking
Solana section for instructions on locking and unlocking the instrument and shutting down.
The Solana must be placed on a dry, stable, and level surface and within the operating conditions specified in
the Technical Specifications section. The location needs to be within reach of an appropriate wall electrical
outlet and other wired connections (e.g., Dymo printer). In addition, it should be placed with enough
surrounding space to perform the following functions:
◼ Easily open and close the tube lid heater to insert or remove the reaction tubes
◼ Easy access to all connection ports on the back of the unit should be considered during placement of
instrument, such as the connector for the power supply, in case of emergency or under abnormal
operating conditions
The measured signal may change due to following:
◼ Changes in ambient light levels
◼ Contamination of the optical parts (Ex. Corrosion from bleach and sample leakage)
◼ Electromagnetic interference
◼ Temperature changes (Ex. Placing under air vents and direct sunlight.)
◼ Mechanical movements (vibrations). (Ex. Centrifuges and large machines.)
Do not place Solana in a location that is subject to vibrations, such as near machines that vibrate, shake,
or spin. Solana is a highly sensitive and precise optical instrument.
Do not place Solana in a location that will subject it to sources of direct airflow, such as an overhead air
vent or an adjacent instrument with a fan vent. Direct airflow can affect Solana Instrument’s ability to maintain
proper temperature.
Do not place Solana near sources of intense light (e.g., direct sunlight). Solana has internal correction for
normal levels of ambient light, but highly intense light falling into the tube holder port may cause significant
interference with the measurement.
Do not place Solana directly next to a microwave or source of electromagnetic interference.
Do not expose Solana to ambient temperatures outside 15°C to 35C.