47
Select Status (In-Patient, Out-Patient, or Proficiency) using the drop-down list.
Note: Sample ID, Age, and Status are required for Virena. If not entered, an error message is generated after
the run is complete.
If needed, enter Study Fields by touching the Open button.
Note: This feature is optional and may be helpful for tracking samples in research studies.
Enter 2-digit numeric values for up to 4 Study Fields.
Touch the Enter key complete.
Touch the button to confirm the Study Fields entries.
Touch the Enter key (Figure 45).
Touch the button to confirm the ID entries for the sample.
Continue with sample entry as described above.
Correcting an Erroneous Manual Sample ID Entry
Touch the tube that contains an entry error.
Enter corrected sample information in the ID field.
Touch the Enter button .
Correcting an Erroneous Barcoded Sample ID Entry
Touch the tube that contains an entry error.
Touch the Delete button to clear the information.
Scan the corrected sample barcode. The corrected ID will automatically populate.
Test in Progress
Once testing has started, the Test in Progress bar will display the test percent completion, and a countdown
timer at the top of the menu will indicate overall test progression (Figure 46).
If needed, to abort the test touch the abort button.
Touch the button to confirm termination of the run.
Discard the tubes after aborting the test(s).
If the aborted run needs to be repeated, please refer to the individual Solana Assay Product Insert to
ensure the original samples are within storage specifications.