Page #61 background imageLoading...
Page #61 background image
61
User Interaction Required (yellow)
Pop-up message
Cause
Solution
Instrument was turned on after
shutdown with the transport
lock activated.
Ensure transport lock is unlocked
and confirm by touching the
green button.
Tube lid heater is open.
Close tube lid to continue.
User selected print option.
No action required.
User selected to install software
update.
Insert USB flash drive (USB stick)
with software update and touch
the USB drive button, or cancel by
touching the red button.
User selected to add a new
Report (printer) Template.
Insert USB flash drive (USB stick)
with new printer template and
confirm by touching green button
or cancel by touching red button.