An example of a printout of a report from adjustment process: 
 
The printout content depends on 
parameters set in group P2 GLP (see point 
8.2 of this user manual). Variables that are 
printed have an attribute set to 1: yes. 
Apart from the content set in this menu 
group, a report contains the following data: 
mass of an adjustment weight stored in 
moisture analyzer memory from last 
completed adjustment process (described as 
Curr:
), mass of an adjustment weight 
determined in current adjustment process 
(described as Adj: (or Cal) and a deviation 
from adjustment process (described as 
), i.e. the difference between the two 
mass values. 
 
8.1.6. Drying test 
 
Parameter <Drying test> enables monitoring correctness of temperature 
indications as recorded by the moisture analyzer’s internal thermometer. 
The testing procedure is carried out using a dedicated set (a thermometer with a 
holder) that is available as additional equipment of a moisture analyzer MAX 
series. (It is the same set that is used for carrying out temperature adjustment 
process). Means of assembling the set is described in the chapter on 
temperature adjustment of this user manual. 
 
Enabling the testing procedure requires assembling the control set inside the 
moisture analyzer’s drying chamber, and entering a menu group <Adjustment> 
in instrument’s parameters. Start the process by following below specified 
instructions. 
Remember to assembly the control thermometer set before starting the testing 
procedure and then close the drying chamber lid.  
 
 
 
Enter menu group P1 Adjustment. 
Set the marker next to a parameter 06 
Drying test 
 
 
 
 
Press RIGHT ARROW navigating key. The 
moisture analyzer opens a window for 
entering the temperature value for carrying 
out the drying test. Use the navigsting keys 
to enter the desired temperature value and 
press ENTER button to accept.