- 24 -
following step of drying chamber adjustment
process. The moisture analyzer continutes
heating the drying chamber for another 8
minutes and simultaneously indicates on its
display the current temperature value. After
the 8 minute time interval the moisture
analyzer’s display indicates the flashing
temperature value. Enter the temperatrure
value as indicated on the external (control)
thermometer. After entering the temperature
value finish the adjustment process by
pressing Print/Enter key.
After completing the process return to the main menu with procedure of saving
changes.
8.1.4. Adjustment test
Adjustment test is a comparison of mass of an adjustment weight value with the
value of current measurement result. The process is aimed at checking the
comparison results, and it does not affect coefficients saved in software settings.
On previewing the test results press ESC key.
Carrying out adjustment test
a) Go to submenu P1 Adjustment.
b) Set the marker next to the Adjustment test function
c) Press RIGHT ARROW navigating key
d) The moisture analyzer automatically carries out the test process
Factory adjustmentname of last carried out adjustment process
Adj (Cal) value of a load currently placed on moisture analyzer’s pan
Curr. mass value of an adjustment weight saved in moisture
analyzer’s memory
Dev.calculated difference between the above two values
e) Return to the weighing mode =, as specified in point 6 of this user manual.
8.1.5. Adjustment report printout
On completing each of adjustment processes the moisture analyzer enables
generating a report from adjustment process. The report can be printed on a
connected printer and sent to a computer and saved in a form of file for future
records.
P1 05 Report printout : 1 : YES report is printed
P1 05 Report printout: 0 : NO report is not printed
Remember that on enabling report printout (set to 1), the report is generated
and sent automatically.