- 33 -
8.5. RS 232 Functions
Moisture analyzer MAX series enables determining communication parameters
with a computer or a printer.
Enter menu group P5 RS-232
8.5.1. RS 232 parameter settings
Use function keys to set required parameters values (see point 7 of this user
manual moving through user menu) which determine correct cooperation between
a moisture analyzer and a computer or a moisture analyzer and a printer:
Parameter
number and
name
Parameter value
Parameter
number and name
Parameter value
01 Baud rate:
0 : 2400;
1 : 4800;
2 : 9600;
3 : 19200
05 Handshake
0 : no;
1 : RTS/CTS;
2 : XON/XOFF
02 Parity
0 : no;
1 : even;
2 : odd
06 Print stable
0 : no;
1 : yes
03 Data bits
1 : 7 bits;
2 : 8 bits
07 Printer type
0 : Standard
1 : Kafka
2 : Epson
3 : PC
4 : PC-RADWAG
04 Stop bits
1 : 1 bit;
2 : 2 bits
08 Paper cut
0 : no
1 : yes
Caution:
In case of parameter set to P5 07 3 : PCthe data sent to a printer is only a line.
The data from the header and the footer is not printed.
For parameter set to P5 07 4 : PC-RADWAGeach printout is proceeded by a
special controlling signs, which enable moisture analyzer cooperation with the
computer software E2R MOISTURE ANALYZER.