- 74 -
14. COOPERATION WITH A PRINTER OR A COMPUTER
Each pressing of < PRINT > key causes sending to a connected printer
or computer a signal corresponding to current mass indication (display
status) with enabled measuring unit. The default baud rate setting of a
balance is 9600 bit/sec. If a peripheral d
evice (a printer or a computer)
requires other baud rate settings then it needs to be changed in the
moisture analyzer’s menu (see point 13 of this user manual).
14.1. Connecting cables
A balance: slot DB 9/F a KAFKA printer: slot WM 560
3 (TxD)
1 (RxD)
5 (GND)
3 (GND)
7- 8 shorted
A balance: slot DB 9/F a computer: slot DB 9/F (if handshake is disabled)
2 (RxD)
3 (TxD)
3 (TxD)
2 (RxD)
5 (GND)
5 (GND)
4 - 6 shorted
4 - 6 shorted
7- 8 shorted
7 - 8 shorted
A balance: slot DB 9/F a computer: slot DB 9/F (if handshake is enabled)
2 (RxD)
3 (TxD)
3 (TxD)
2 (RxD)
4 (DTR)
4
5 (GND)
5 (GND)
6 (Tara)
6
7 (RTS)
7 (CTS)
8 (CTS)
8 (RTS)
9 (Print)
9
A balance: slot DB 9/F a computer: slot DB 25/F
2 (RxD)
2 (TxD)
3 (TxD)
3 (RxD)
5 (GND)
7 (GND)
7 - 8 shorted
4 - 5 shorted
6 - 20 shorted

Table of Contents