movement) it is recommended to set the filter value to slow or very slow (set value
of parameter 01 Filter to 4 or 5). The efficiency of filter operation differs in relation
to the measuring range. The filter operates with lower accuracy while the mass
indication is quickly increasing after placing a load on the pan, Filter accuracy is
increased when weighed mass is within filter’s set operation range (parameter:
filter operation range is available only in balance’s service menu, and it is
inaccessible for the user).
8.4.3. Autozero function
In order to ensure moisture analyzers’s precise mass indication, “AUTOZERO”
software parameter has been introduced. The application of this function is
automatic control and correction of zero indication.
The parameters determining the range and time of autozero operation are
set by the manufacturer and they are not accessible for the user.
If AUTOZERO function is enabled, then each weighing process starts from
precise zero point. Disabling AUTOZERO operation may result in result instability
while the pan is not loaded with a mass to be weighed. There are, however, some
cases when this function can be a disturbing factor of measuring process; for
instance very slow placing of a load on the weighing pan (e.g. load pouring).
Enabling and disabling AUTOZERO operation is carried out in parameter P4 02.
8.4.4. Previewing temperature value inside the drying chamber